Primary Contact: Name | Yngve Cerenius |
Owner/Operator | Max-Lab at Lund University |
Status | Decommissioned |
%Time Available (general use) | 100 |
Experiments | MAD, SAXS, POWDER |
Experiments Details | The beamline has a single crystal monochromator which provides a high photon flux at the sample while sacrificing easy tunability and high energy resolution. The experimental area is designed to have high flexibility for arranging different experimental set-ups and for easy exchange between different area and position sensitive detectors. |
Source Details | 13-period, 1.8 T, multipole-wiggle |
Spot Size (mm): Height | 0.3 |
Spot Size (mm): Width | 0.3 |
Monochromator | Triangular single crystal with horizontal diffraction plane |
Crystal Type | Si (111) |
Mirrors | Vertically focusing Pt coated Si Mirror - bendable for horizontal focusing |
Flux(PHOTONS/SEC) | 1 x 10^11 |
Flux Details | photons/(sec 0.1% band width).calculated -> 1011 photons/sec for wavelenghts > 1 Å, behind slits 0.3 x 0.3 mm2. The additional 3 slit collimation of the incoming X-ray beam reduces the intensity to 1010 photons/sec for the SAXS experiments. |
Wavelength Range (Å) | 0.81 - 1.55 |
Energy Range (keV) | 8.0 - 14.6 |
1. Detector: Type | CCD |
1. Detector: Model | BRUKER SMART 1000 |
1. Detector: Details | detector for macromolecular crystallography mounted on a Mardtb baseplate. It has also successfully been used for powder diffraction experiments together with the Fit2D software. |
1. Detector: Activated for Service | info currently not available |
1. Detector: Removed from Service | |
2. Detector: Type | IMAGE FOIL |
2. Detector: Model | HUBER 670 |
2. Detector: Details | with both a flat specimen and capillary holder as well as furnace is used for powder diffraction experiments. |
2. Detector: Activated for Service | info currently not available |
2. Detector: Removed from Service | |
3. Detector: Type | CCD |
3. Detector: Model | BRUKER SMART 1000 |
3. Detector: Details | as back-up - equipped with an extended base with a maximum crystal to detector distance of 770 mm. |
3. Detector: Activated for Service | info currently not available |
3. Detector: Removed from Service | |
4. Detector: Type | CCD |
4. Detector: Model | BRUKER SMART 1000 |
4. Detector: Details | ith 90 mm diameter active area and 1024*10124 pixels is mounted on a three circle fixed chi goniometer. The system is equipped with a rotary shutter, collimators down to 50 micron and software packages for on-site analysis of both powder and single crystal data. |
4. Detector: Activated for Service | info currently not available |
4. Detector: Removed from Service | |
Cryo Capability | Oxford instruments cryoject and cryostream |
Lab Facilities | wax melter, liquid nitrogen and microscopes |
Software: Name | Automar, HKL2000,Mosflm, XDS, CCP4, O, USF, X-PLOR, Bruker package for SMART 1K |
Software: Classification | |
Hardware | mLinux stations and Irix station for processing the data acquired on the marccd detector, one Windows machine in the computer room for processing the data from the SMART and Huber detectors |
Backup Services | DAT tapes, CD/DVD, firewire |