| Primary Contact: Name | Seiki Baba |
| Secondary Contact: Name | Takashi Kumasaka |
| Owner/Operator | Japan Synchrotron Radiation Research Institute (JASRI) |
| Status | Operational |
| %Time Available (general use) | 80 |
| Experiments | MAD, SAD, MONOCHROMATIC |
| Services | Data collection, Mail-in |
| Source Type | BENDING MAGNET |
| Spot Size (mm): Height | 0.053-0.180 |
| Spot Size (mm): Width | 0.087 |
| Monochromator | Fixed exit double crystal monochromator |
| Crystal Type | Si (111) |
| Mirrors | Vertically bent cylinderical mirror coated in Rhodium |
| Flux(PHOTONS/SEC) | 9.3 x 10^10 |
| Flux Details | Wavelength 1.0 A, beam size 0.087x0.180 mm |
| Wavelength Range (Å) | 0.75 - 1.90 |
| Energy Range (keV) | 6.5 - 16.5 |
| Goniometer | Horizontal axis goniometer having motor-driven x, y, z stage |
| Sample Handling: Robotics | (SPACE) Spring-8 Precise Automatic Cryo-Sample Exchanger |
| Robotics: More Info | Sample Magazine: 2 Uni-pucks or 2 SPACE trays |
| Detector: Type | CCD |
| Detector: Model | ADSC Quantum 315r |
| Detector: Activated for Service | info currently not available |
| Detector: Removed from Service | |
| Cryo Capability | Cryostat : temperature control range 80 K - 350 K |
| Lab Facilities | Biology and chemistry preparation laboratories |
| Software: Name | CrystalClear (RIGAKU/MSC), HKL2000 (HKL Research, Inc.) |
| Software: Classification | |
| Hardware | UV-Vis microspectrophotometer |