| Primary Contact: Name | Go Ueno |
| Owner/Operator | RIKEN Harima Institute |
| Status | operational |
| Experiments | MAD |
| Source Type | bending magnet |
| Monochromator | Fixed exit Si double crystal monochromator |
| Crystal Type | Si (111) |
| Mirrors | vertically bent two dimensional focusing mirror coated in Rhodium |
| Flux(PHOTONS/SEC) | 1 x 10^11 |
| Flux Details | at 12 keV monochromatic |
| Wavelength Range (Å) | 0.729 - 2.066 |
| Energy Range (keV) | 6 - 17 |
| Goniometer | micro kappa |
| Sample Handling: Robotics | (SPACE) Spring-8 Precise Automatic Cryo-Sample Exchanger |
| 1. Detector: Type | CCD |
| 1. Detector: Model | RIGAKU JUPITER 210 |
| 1. Detector: Activated for Service | info currently not available |
| 1. Detector: Removed from Service | |
| 2. Detector: Type | IMAGE PLATE |
| 2. Detector: Model | RIGAKU R-AXIS V |
| 2. Detector: Activated for Service | info currently not available |
| 2. Detector: Removed from Service | |
| Cryo Capability | Continuous cold stream of nitrogen generated from atmosphere. The temperature range is 90 to 300K. |
| Lab Facilities | Biology and chemistry preparation laboratories |
| Software: Name | CrystalClear (RIGAKU/MSC), HKL2000 (HKL Research, Inc.) |
| Software: Classification | |